Ceramic Fracture Following Cervical Disc Arthroplasty

Ngoc Quyen Nguyen, Dinesh Kafle, Jacob M. Buchowski, Kunwoo Park, Bong‐Soon Chang, Choon‐Ki Lee, Jin S. Yeom

Journal of Bone and Joint Surgery · 2011 · 17 citations · 9 references

Concepts

Abstract

Although ceramic fractures have been reported following
\ntotal hip arthroplasty1-6, with an incidence ranging from
\n0.004% to 1.4%1-3, no cases of ceramic fracture following
\ncervical disc arthroplasty have been reported, to our
\nknowledge. We present the case of a patient with cervical radiculopathy
\nwho underwent total disc replacement at C5-C6
\nand C6-C7, which was complicated by the development of
\nrecurrent symptoms approximately one month after the index
\nprocedure. At the time of the revision surgery, the ceramic
\ninsert at the C6-C7 level was found to be fractured on its convex
\ncranial side. The implant used was a semiconstrained ceramicon-
\nceramic prosthesis (Discocerv Cervidisc Evolution; Scientx,
\nGuyancourt, France). Since the introduction of this
\nprosthesis in April 20067, more than 2000 prostheses have
\nbeen implanted. The patient was informed that data concerning
\nthe case would be submitted for publication, and he
\ngave his consent.