The International Journal of Artificial Organs · 2007 · 15 citations · 21 references
EngineeringCardiologyPhysicsComputational Fluid DynamicsFluid DynamicsFluid MechanicsCardiovascular BiomechanicsExperimental Fluid MechanicsFlow MeasurementFlow ModelingHemodynamicsHemodynamics ModelingCfd SimulationCfd ValidationComputational Fluid Dynamics ValidationFluid - Structure Interaction ModelingFluid-structure Interaction ModelingParticle Image VelocimetryPulsatile FlowShear Stress
The aim of this study was to validate the 2D computational fluid dynamics (CFD) results of a moving heart valve based on a fluid-structure interaction (FSI) algorithm with experimental measurements. Firstly, a pulsatile laminar flow through a monoleaflet valve model with a stiff leaflet was visualized by means of Particle Image Velocimetry (PIV). The inflow data sets were applied to a CFD simulation including blood-leaflet interaction. The measurement section with a fixed leaflet was enclosed into a standard mock loop in series with a Harvard Apparatus Pulsatile Blood Pump, a compliance chamber and a reservoir. Standard 2D PIV measurements were made at a frequency of 60 bpm. Average velocity magnitude results of 36 phase-locked measurements were evaluated at every 10 degrees of the pump cycle. For the CFD flow simulation, a commercially available package from Fluent Inc. was used in combination with inhouse developed FSI code based on the Arbitrary Lagrangian-Eulerian (ALE) method. Then the CFD code was applied to the leaflet to quantify the shear stress on it. Generally, the CFD results are in agreement with the PIV evaluated data in major flow regions, thereby validating the FSI simulation of a monoleaflet valve with a flexible leaflet. The applicability of the new CFD code for quantifying the shear stress on a flexible leaflet is thus demonstrated.
21
Functional Living Trileaflet Heart Valves Grown In Vitro
Simon P. Hoerstrup, Ralf Sodian, Sabine Daebritz et al. · Circulation · 2000 · 591 citations