Physics in Medicine and Biology · 2015 · 62 citations · 25 references
Natural SciencesPhysicsRadiologyRadiation DetectionRadiation DosimetryRadiation MeasurementComputed Tomography PhysicsElectron CountingCt Number To Electron Density CalibrationDual-energy Computed TomographyDual-energy ImagingEffective Atomic NumberElectron DensityElectron Density AnalysisElectron Density DistributionsKarhunen-loève Expansion
The purpose of this work is to present a new method to extract the electron density ([Formula: see text]) and the effective atomic number (Z eff) from dual-energy CT images, based on a Karhunen-Loeve expansion (KLE) of the atomic cross section per electron. This method was used to calibrate a Siemens Definition CT using the CIRS phantom. The predicted electron density and effective atomic number using 80 kVp and 140 kVp were compared with a calibration phantom and an independent set of samples. The mean absolute deviations between the theoretical and calculated values for all the samples were 1.7 % ± 0.1 % for [Formula: see text] and 4.1 % ± 0.3 % for Z eff. Finally, these results were compared with other stoichiometric method. The application of the KLE to represent the atomic cross section per electron is a promising method for calculating [Formula: see text] and Z eff using dual-energy CT images.
25
WinXCom—a program for calculating X-ray attenuation coefficients
L. Gerward, Nicholas R. Guilbert, Kasper Jensen et al. · Radiation Physics and Chemistry · 2004 · 1.1K citations
Atomic weights of the elements 2011 (IUPAC Technical Report)
Michael E. Wieser, N.E. Holden, Tyler B. Coplen et al. · Pure and Applied Chemistry · 2013 · 416 citations · Full text