Publication | Closed Access
A new method to measure electron density and effective atomic number using dual-energy CT images
61
Citations
25
References
2015
Year
EngineeringNuclear PhysicsMicroscopyNuclear DataElectron DiffractionDual-source CtEffective Atomic NumberElectron PhysicElectron MicroscopyElectron SpectroscopyNuclear MaterialsPhoton-counting Computed TomographyNuclear MedicineRadiologyElectron DensityAccelerator Mass SpectrometryMedical ImagingPhysicsMonte-carlo ModellingPredicted Electron DensityAtomic PhysicsExperimental Nuclear PhysicsNatural SciencesSpectroscopyBiomedical ImagingDual-energy Ct ImagesElectron MicroscopeCalibration Phantom
The purpose of this work is to present a new method to extract the electron density ([Formula: see text]) and the effective atomic number (Z eff) from dual-energy CT images, based on a Karhunen-Loeve expansion (KLE) of the atomic cross section per electron. This method was used to calibrate a Siemens Definition CT using the CIRS phantom. The predicted electron density and effective atomic number using 80 kVp and 140 kVp were compared with a calibration phantom and an independent set of samples. The mean absolute deviations between the theoretical and calculated values for all the samples were 1.7 % ± 0.1 % for [Formula: see text] and 4.1 % ± 0.3 % for Z eff. Finally, these results were compared with other stoichiometric method. The application of the KLE to represent the atomic cross section per electron is a promising method for calculating [Formula: see text] and Z eff using dual-energy CT images.