Journal of Clinical Microbiology · 2015 · 24 citations · 12 references
Veterinary MedicineInfectious DiseasesClinical ImmunologyDiagnostic MicrobiologyMolecular DiagnosticsAntigen Detection AssaysHiv DiagnosisHiv TestingPoint-of-care Diagnostic DevicesRapid Diagnostic TestsSexually Transmitted Infection DiagnosticsClinical Laboratory EvaluationHiv Antibody DetectionHiv Serologic TestingLaboratory Diagnosis Of HivLaboratory Performance AssessmentLaboratory Test EvaluationLaboratory TestingPoint-of-care Immunodiagnostics
We assessed the laboratory performance of the Chembio dual-path platform HIV-syphilis rapid immunodiagnostic test and electronic reader for detection of HIV and Treponema pallidum antibodies in 450 previously characterized serum specimens. For visual or electronic reader HIV antibody detection, the sensitivity was 100% and the specificity was 98.7%. For visual T. pallidum antibody detection, the test sensitivity was 94.7% and the specificity was 100.0%; with the electronic reader, the sensitivity was 94.7% and the specificity was 99.7%.
12
Claire Bristow, Yaw Adu‐Sarkodie, Raphael Ondondo et al. · Open Forum Infectious Diseases · 2014 · 47 citations · Full text
Melvy Ormaeche, Álvaro Whittembury, Mónica Pun et al. · International Journal of Infectious Diseases · 2012 · 42 citations · Full text