Effect of Wear and Crevice on the Corrosion Resistance of Overlapped Stents

Christine Trépanier, Xiao-Yan Gong, Tom Ditter, Alan R. Pelton, Yasmin Neely, R.B. Grishaber

2008 · 14 citations · 5 references

Concepts

Abstract

This study assessed the effect of wear damage and crevice corrosion on Nitinol and stainless steel stents. Wear between the Nitinol or stainless steel stents was generated through in vitro axial fatigue tests of overlapped stents. SEM investigation of the stents after the fatigue test characterized the extent of wear damage on both Nitinol and stainless steel stents. Potentiodynamic polarization tests based on ASTM F2129 showed that both Nitinol and stainless steel are characterized by a decrease in the breakdown potential from about 900 mV vs SCE pre- fatigue test to about 550-650 mV vs SCE post-fatigue test. These results indicate that both groups of material are affected similarly by wear damage. In a separate study, potentiostatic polarization testing (ASTM F746) was used to evaluate the effect of crevice corrosion created by gaps between two overlapping Nitinol or stainless steel stents. The results from this study showed that both materials were able to repassivate crevice corrosion damage at 800 mV vs SCE.