Fixation of tibial components of knee prostheses.

P. S. Walker, D Greene, Donald T. Reilly, J Thatcher, M. Ben-Dov, F C Ewald

Journal of Bone and Joint Surgery · 1981 · 135 citations · 0 references

Concepts

Abstract

Twelve different tibial components were compared for fixation. The components included compartmental, anteriorly joined, posterior-cruciate retaining, and one-place with one, two, or three fixation posts; all-plastic or with a metal tray. The apparatus applied compressive load with anterior-posterior force, rotational torque, or varus-valgus moment. The relative deflections, both compressive and distractive, were measured between the component and the bone. The least deflections occurred with one-piece metal-tray components with one or two posts and with one-piece plastic components with a post on each side. Compartmental components deflected the most.