Journal of Bone and Joint Surgery · 1981 · 135 citations · 0 references
MedicineOrthopaedic SurgeryOrthopedic SurgeryOrthopedic BiomechanicsOrthopedicsProstheticsArthroplasty BiomechanicsFixation BiomechanicsKnee ArthroplastyKnee BiomechanicsProsthetic BiomechanicsCompartmental BiomechanicsCompressive LoadsKnee Joint LoadingTibial Component DesignTibial Component FixationTotal Knee Arthroplasty Biomechanics
Twelve different tibial components were compared for fixation. The components included compartmental, anteriorly joined, posterior-cruciate retaining, and one-place with one, two, or three fixation posts; all-plastic or with a metal tray. The apparatus applied compressive load with anterior-posterior force, rotational torque, or varus-valgus moment. The relative deflections, both compressive and distractive, were measured between the component and the bone. The least deflections occurred with one-piece metal-tray components with one or two posts and with one-piece plastic components with a post on each side. Compartmental components deflected the most.