Reviewing Polycrystalline Mercuric Iodide X-Ray Detectors

M. Schieber, Asaf Zuck, H. Gilboa, George Zentai

IEEE Transactions on Nuclear Science · 2006 · 52 citations · 9 references

Concepts

Abstract

Polycrystalline films of HgI <sub xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">2</sub> , which are used as nuclear radiation detectors, are fabricated by two different methods, either by Physical Vapor Deposition (PVD) or glued with a polymeric binder called also "Particle in Binder" (PIB). Both types of PVD or PIB Detectors are deposited directly on Thin Film Transistor TFT technology imagers and can be used for X-ray imaging. The present review is concentrated on PVD- Polycrystalline-HgI <sub xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">2</sub> . However, the main review body is a summary of the work performed by RTR, HU, and Varian. Works by other groups are mentioned briefly. The PVD process is explained by reviewing X-ray Diffraction (XRD) and Scanning Electron Microscope (SEM) data. The charge transport properties of the PVD-Polycrystalline HgI <sub xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">2</sub> detectors as they improved with R&D efforts are shown. The response of these PVD detectors to <sup xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">241</sup> Am and <sup xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">57</sup> Co gamma rays and their X-ray imaging properties such as dark currents, sensitivity, imaging resolution and Detector Quantum Efficiency (DQE) are reviewed