Publication | Closed Access
Investigation of the performance of the General Electric ADVANCE positron emission tomograph in 3D mode
143
Citations
12
References
1996
Year
Computed TomographyEngineeringReconstruction TechniquePhysicsMedical ImagingHoffman Phantom3D ImagingCt ScanSurgeryComputational ElectromagneticsGe Advance 3DInstrumentationSynchrotron RadiationRadiation ImagingNema PhantomsRadiologyHealth Sciences
Performance measurements of the General Electric ADVANCE Positron Emission Tomograph operating with the septa retracted (3D mode) were made. All reconstructions were performed with the GE ADVANCE 3D package. Performance tests were carried out with: the NEMA phantoms; a 3D Hoffman phantom; a Data Spectrum torso phantom with lung and cardiac inserts; and the "Utah" 3D evaluation phantom. Data collected included: transaxial and axial resolution, uniformity, recovery coefficients, count rate performance, dead time accuracy, and effect of scatter correction.