Magnetic Resonance in Medicine · 1987 · 18 citations · 7 references
Natural SciencesBiophysicsRadiologyDiagnostic ImagingDiagnostic RadiologySpectroscopyMagnetic Resonance ImagingMagnetic Resonance Imaging PhysicsMagnetic Resonance Pulse SequencesMagnetic Resonance Signal FormationMagnetic Resonance SpectroscopyMedical ImagingChemical Shift ImagingField HomogeneityMagnetic Field InhomogeneityMagnetic Resonance Imaging AcquisitionMagnetic Resonance Spectroscopic ImagingPulse Sequence DesignPulse Sequences
A new echo-time-encoded chemical-shift imaging technique applicable to dual-peak spectroscopic imaging with large magnetic field inhomogeneity is proposed and studied. The basic concept and its applications to modest field homogeneity (approximately 3.0 ppm) as well as to relatively large field inhomogeneity (approximately 10.5 ppm) are discussed. Actual pulse sequences are given and some experimental results on human volunteers obtained with a 2.0-T KAIS NMR system are also presented.
7