IEEE Transactions on Rehabilitation Engineering · 1993 · 48 citations · 14 references
MedicineBiomedical EngineeringOrthopedic SurgeryAmputee RehabilitationBiomechanicsProstheticsBiomechanical InstrumentationLower-limb ProstheticsProsthetic BiomechanicsProsthetic RehabilitationSocket DesignStrain SensorsInstrumentation Error AnalysisInstrumentation Error SourcesLimb - Socket Interface MechanicsProsthetic Socket BiomechanicsProsthetic Socket InterfaceResidual Limb BiomechanicsShear Stress At The Prosthetic InterfaceWalking
Transducers were developed to measure stresses in three orthogonal directions applied on a below-knee amputee residual limb within a prosthetic socket during walking. Strain-gage measurement techniques were used for all three orthogonal directions. Instrumentation error for four transducers averaged less than 4.2% full-scale output (FSO) for the normal direction and less than 0.9% FSO for shear directions. Measurements from amputee subjects collected during walking trials are presented.< <ETX xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">></ETX>
14
EXPERIMENTAL FRICTION BLISTERS.
P.F.D. Naylor · British Journal of Dermatology · 1955 · 146 citations