Measurement of stresses in three orthogonal directions at the residual limb-prosthetic socket interface

Joan E. Sanders, Colin Daly

IEEE Transactions on Rehabilitation Engineering · 1993 · 48 citations · 14 references

Concepts

Abstract

Transducers were developed to measure stresses in three orthogonal directions applied on a below-knee amputee residual limb within a prosthetic socket during walking. Strain-gage measurement techniques were used for all three orthogonal directions. Instrumentation error for four transducers averaged less than 4.2% full-scale output (FSO) for the normal direction and less than 0.9% FSO for shear directions. Measurements from amputee subjects collected during walking trials are presented.< <ETX xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">&gt;</ETX>