Method of measuring 'shearing' forces on the sole of the foot

J.W. Tappin, Jennifer Pollard, E A Beckett

Clinical Physics and Physiological Measurement · 1980 · 72 citations · 1 references

Concepts

Abstract

A system is described which enables recordings from metatarsal heads and the toe to be made simultaneously while the subject is walking bare-foot, and with a variety of footware to test their effect in modifying and redistributing the shear forces. The system involves placing a centre-tapped magneto resistor in a bridge configuration, with a magnet placed centrally above it. By applying an elastic self-centring force to the magnet, the degree of movement is made proportional to the shearing force applied.

References

1