IEEE Transactions on Microwave Theory and Techniques · 2004 · 468 citations · 21 references
EngineeringBioengineeringBiomedical EngineeringAntennasBiocompatibilityBiomaterialsAntenna DesignImplantable Biomedical DevicesImplantable MicrosystemsMedical Device DesignMicrostrip AntennasMicrowave AntennasBiomedical ImplantsFinite Difference Time Domain ModelingFinite Difference Time Domain SimulationImplantable Antenna DesignImplantable AntennasMicrostrip Antenna DesignMicrostrip Patch AntennasPatch Antennas
The paper aims to design a microstrip patch antenna for communication with medical implants in the 402‑405 MHz Medical Implant Communications Services band. Finite‑difference time‑domain simulations were used to evaluate and optimize antenna parameters—including shape, size, feed and ground point locations, substrate and superstrate materials and thicknesses—against measured results. The results provide design guidance for implantable microstrip antennas in the 402‑405 MHz band.
The objective of this paper is to design a microstrip patch antenna for communication with medical implants in the 402-405-MHz Medical Implant Communications Services band. Microstrip antenna design parameters are evaluated using the finite-difference time-domain method, and are compared to measured results. The effects of shape, length, size, location of feed point and ground point, substrate and superstrate materials, and their thicknesses are evaluated. An extensive study of the performance of the antennas to changes in these parameters was undertaken. The results of this paper provide guidance in the design of implantable microstrip antennas.
21