An Emittance Scanner for Intense Low-Energy Ion Beams

P.W. Allison, Joseph D. Sherman, D. B. Holtkamp

IEEE Transactions on Nuclear Science · 1983 · 143 citations · 0 references

DOIFull text

Open access

Concepts

TL;DR

An emittance scanner was developed for low‑energy H⁻ beams to provide ±0.5 mrad angular resolution, minimal space‑charge errors, and a compact, simple design. The 10‑cm analyzer uses two slits and ±500‑V electric deflection plates; as it scans, the front slit selects a thin beam ribbon, and the ion angle—proportional to the ramp voltage—determines the current through the rear slit, yielding phase‑space density that is computer‑analyzed for rms emittance and density contours. Measured emittance and phase‑space density contours agree well with calculations from a prepared collimated phase space.

Abstract

An emittance scanner has been developed for use with low-energy H- ion beams to satisfy the following requirements: (1) angular resolution of ± 1/2 mrad, (2) small errors from beam space charge, and (3) compact and simple design. The scanner consists of a 10-cm-long analyzer containing two slits and a pair of electric deflection plates driven by a ±500-V linear ramp generator. As the analyzer is mechanically driven across the beam, the front slit passes a thin ribbon of beam through the plates. The ion transit time is short compared with the ramp speed; therefore, the initial angle of the ions that pass through the rear slit is proportional to the instantaneous ramp voltage. The current through the rear slit then is proportional to the phase-space density d2i/dxdx. The data are computer-analyzed to give, for example, rms emittance and phase-space density contours. Comparison of measured data with those calculated from a prepared (collimated) phase space is in good agreement.