Nosocomial Transmission of Imipenem-Resistant <i>Pseudomonas Aeruginosa</i> Following Bronchoscopy Associated With Improper Connection to the STERIS SYSTEM 1 Processor
89
Citations
34
References
2001
Year
The similarity of diverse connectors and limited training by the manufacturer regarding AER for bronchoscopes were the two factors responsible for the outbreak. Appropriate connections were implemented, and there was no further bronchoscope contamination. We suggest active surveillance of all bronchoscopy specimen cultures, standardization of connectors of various scopes and automated processors, and systematic education of staff by manufacturers with periodic on-site observation.