2003 · 35 citations · 2 references
Upright PosturePortable Data LoggerAccelerometerWearable TechnologySpine DeformityPostureHuman MonitoringOrthopaedic SurgeryLumbar SpineKinesiologyApplied PhysiologySpinal PostureKinematicsHealth SciencesInertial SensorsAssistive TechnologyPosture Monitoring SystemRehabilitationSagittal PlanePhysical TherapyHuman MovementMedicine
A new method of monitoring spinal posture in the sagittal plane is presented. Posture is measured during activities of daily living utilizing accelerometer inclinations of six points along a subject's spine. The over sampled, low frequency filtered data can be monitored and collected over extended periods (up to four weeks) utilizing the inexpensive, low-power, portable data logger developed in the course of this research. The data logger utilizes 7 processors to collect and store data onto an MS-DOS formatted CompactFlash card for later analysis. Two trials were done: regimented postures across a one hour period and unregimented activities during a five hour period. Data was then visualized graphically and compared to noted postures and activities. The system was designed to be accurate to within /spl plusmn/0.5/spl deg/. Measured precision was found to be within /spl plusmn/0.39/spl deg/. This system will be useful for monitoring changes in back curvature resulting from surgery or treatment.
2