IEEE Transactions on Ultrasonics Ferroelectrics and Frequency Control · 2007 · 43 citations · 13 references
MedicineBiomedical EngineeringRadiologyDiagnostic ImagingUltrasonographyUltrasoundContrast AgentsDiagnostic UltrasonographyMedical ImagingUltrasonicsUltrasound ImagingUltrasound PhysicsAmplitude ModulationBandwidthContrast RatioMicrobubble Contrast ImagingNonlinear FilteringPulse CompressionPulse Inversion ImagingUltrasound Contrast Agents
We present a technique that uses Golay phase encoding, pulse inversion, and amplitude modulation (GPIAM) for microbubble contrast agent imaging with ultrasound. This technique improves the contrast-to-tissue ratio (CTR) by increasing the time-bandwidth product of the insonating waveforms. A nonlinear pulse compression algorithm is used to compress the signal energy upon receive. A 6.5-dB improvement in CTR was observed using an 8-chip GPIAM sequence compared to a conventional pulse-inversion amplitude-modulation sequence. The CTR improvement comes at the cost of a reduction in frame rate: GPIAM coding uses four input pulses whereas most contrast imaging sequences require two or three pulses. Our results showed that the microbubble response can be phase encoded and subsequently compressed using a nonlinear matched-filtering algorithm, in order to enhance the signal from the contrast agent, while maintaining resolution and suppressing the tissue signal.
13
M. Golay · IEEE Transactions on Information Theory · 1961 · 1.5K citations
The Fourier Transform and its Applications
K.W. Cattermole · Electronics and Power · 1965 · 1.4K citations
David Hope Simpson, Chien Ting Chin, Peter N. Burns · IEEE Transactions on Ultrasonics Ferroelectrics and Frequency Control · 1999 · 775 citations
Pulse Inversion Imaging of Liver Blood Flow
Peter N. Burns, Stephanie R. Wilson, David Hope Simpson · Investigative Radiology · 2000 · 355 citations