IEEE Transactions on Medical Imaging · 1991 · 58 citations · 11 references
Natural SciencesBiomedical EngineeringRadiologyBiomedical ImagingDiagnostic ImagingDiagnostic RadiologyImagingMagnetic Resonance ImagingMagnetic Resonance Imaging PhysicsMedical ImagingMedical Imaging PhysicsNeuroimaging PhysicsArtifacts In Magnetic Resonance ImagingDiffusion EquationDiffusion Equation ModelsEddy Current CorrectionEddy CurrentsFilteringInverse ImagingNuclear Magnetic Resonance Studies
Reduction of eddy currents by a temporal compensation of the input current waveform to the gradient coil is studied with an analytic solution. The technique is the inverse filtering of the eddy-current affected field response, which is calculated from the diffusion equation. The limitation of the temporal compensation due to the spatially variant eddy currents is also investigated for whole-body diagnostic imaging systems and small-bore nuclear magnetic resonance (NMR) microscopy systems. Within a limited imaging volume of less than 60% of the gradient coil diameter, most of eddy-current problems can be solved by the technique.
11
The effects of random directional distributed flow in nuclear magnetic resonance imaging
C.B. Ahn, S. Y. Lee, O. Nalcioǧlu et al. · Medical Physics · 1987 · 99 citations