Multi‐residue liquid chromatography/tandem mass spectrometric analysis of beta‐agonists in urine using molecular imprinted polymers

Nathalie Van Hoof, D. Courtheyn, Jean‐Philippe Antignac, M Van de Wiele, Sofie Poelmans, H. Noppe, Hubert De Brabander

Rapid Communications in Mass Spectrometry · 2005 · 56 citations · 12 references

Concepts

Abstract

Ion suppression, a matrix effect that affects quantitative mass spectrometry, is one of the main problems encountered in liquid chromatography/tandem mass spectrometry. Two different clean-up steps for the multi-residue analysis of beta-agonists in urine were evaluated with respect to minimisation of ion suppression, namely, a mixed-phase solid phase extraction (SPE) column, i.e., clean screen Dau (CSD), and a molecular imprinted polymer (MIP) SPE column. Ion suppression experiments revealed that CSD sample clean-up can lead to false negative results for some beta-agonists, and that clean-up using MIP columns is more selective for beta-agonists than the use of CSD columns.