Journal of ASTM International · 2006 · 16 citations · 12 references
MedicineOrthopaedic SurgeryOrthopedic SurgeryJoint Replacement SurgeryOrthopaedic BiomaterialsOrthopedicsArthroplasty BiomechanicsImplant BiomechanicsImplant DesignImplant MaterialsJoint Replacement MaterialsWear ResistanceAdhesive WearCrackingCracking BehaviorDamage MechanismsMaterial Wear AnalysisOrthopaedic ImplantsProsthesis Wear MechanismsSurface Damage
Abstract The purpose of this study was to evaluate the wear and surface damage mechanisms of polyethylene in retrieved CHARITÉ total disc replacements. The retrieved implants showed evidence of adhesive/abrasive wear mechanisms in the central domed region of the implants. In addition, there was also evidence of macroscopic rim damage, including radial cracking, plastic deformation, and third body damage. The retrieved total disc replacements (TDRs) displayed surface damage observed previously in both hip and knee replacements. The information from this study will be useful for engineers seeking to adequately replicate long-term, clinically relevant damage modes of TDRs using in vitro testing methods.