A new linear ion trap mass spectrometer

James W. Hager

Rapid Communications in Mass Spectrometry · 2002 · 463 citations · 17 references

Concepts

TL;DR

The study characterizes mass‑selective axial ion ejection from a linear quadrupole ion trap operated with an auxiliary quadrupole field. The authors employ the ion path of a triple‑quadrupole mass spectrometer, using either the q2 collision cell or the final mass analysis quadrupole as a linear trap, and describe two hybrid RF/DC quadrupole‑linear ion‑trap instruments that enable high‑sensitivity product‑ion scanning while preserving conventional triple‑quadrupole scan modes. Efficient coupling of radial and axial motion in the exit fringing fields allows up to 20 % of trapped ions to be ejected, yielding high‑sensitivity mass spectra; space‑charge distortions are markedly reduced at low pressure, though RF heating from high ejection q‑values and low collision frequencies may also play a role. © 2002 John Wiley & Sons, Ltd.

Abstract

Abstract Characteristics of mass selective axial ion ejection from a linear quadrupole ion trap in the presence of an auxiliary quadrupole field are described. Ion ejection is shown to occur through coupling of radial and axial motion in the exit fringing fields of the linear ion trap. The coupling is efficient and can result in extraction of as much as 20% of the trapped ions. This, together with the very high trapping efficiencies, can yield high sensitivity mass spectral responses. The experimental apparatus is based on the ion path of a triple quadrupole mass spectrometer allowing either the q2 collision cell or the final mass analysis quadrupole to be used as the linear trap. Space charge induced distortions of the mass resolved features while using the pressurized q2 linear ion trap occur at approximately the same ion density as reported for conventional three‐dimensional ion traps. These distortions are, however, much reduced for the lower pressure linear trap possibly owing to the proposed axial ejection mechanism that leads to ion ejection only for ions of considerable radial amplitude. RF heating due to the high ejection q ‐value and the low collision frequency may also contribute. Two hybrid RF/DC quadrupole‐linear ion trap instruments are described that provide high sensitivity product ion scanning while operated in the linear ion trap mode while also retaining all conventional triple quadrupole scan modes such as precursor ion and neutral loss scan modes. Copyright © 2002 John Wiley & Sons, Ltd.

References

17