A Mole for Measuring Pulsed Superconducting Magnets

Pierre Schnizer, Helge Kiesewetter, Thomas Mack, Thomas Knapp, F. Klos, Maximilian Manderla, Stewart E. Rauch, Martin Schönecker, R. Werkmann

IEEE Transactions on Applied Superconductivity · 2008 · 13 citations · 3 references

Concepts

Abstract

The facility for antiproton and ion research (FAIR) at Gesellschaft fur Schwerionenforschung (GSI) will build two superconducting synchrotrons requiring fields of up to up to 2 T at 4 T/s (SIS 100) and 6 Tesla at 1 T/s (SIS 300). The magnets of these rings will be measured using a mole which is currently under development. The main component of the field must be measured with a precision of 250 ppm and the higher order harmonics with a precision better than 100 ppm for DC and pulsed operation. This paper is focused on the pulse measurement using the "step by step" method.

References

3