Analytical Letters · 1995 · 11 citations · 10 references
PharmacyPharmaceutical AnalysisPharmacologyAnalytical ToxicologyChromatographic AnalysisClinical PharmacologyBioanalytical Method ValidationChromatographic Analysis Of PharmaceuticalsChromatographyDrug DispositionInstrumental AnalysisLiquid ChromatographyAssay DeterminationBioanalytical ChromatographyConcentrationHigh-performance Liquid ChromatographyInternal Standard AnalysisMethod Detection LimitReversed-phase ChromatographyReversed-phase Liquid Chromatography
Abstract A simple and rapid high-performance liquid chromatographic (HPLC) method for the determination of amiodarone (AD) in plasma and tissues was developed. The method involved deproteinization of plasma or homogenized tissue with acetonitrile containing an internal standard (N-Cetylpyridinium chloride) followed by reversed phase chromatography using μ bondapack C18 column (10μm) with a mobile phase consisting of acetonitrile - methanol - sodium dihydrogen phosphate buffer (70:10:20%, v/v), the pH adjusted to 4.0 and pumped at flow rate of 1.0 ml/min. The column effluent was monitored at 242 nm. A linear relationship was obtained between peak height ratios (drug to internal standard) versus drug levels over the concentration range of 50–750 ng/ml. The detection limit of AD in plasma and tissues by this method was 20 ng/ml.
10
A JOURNAL OF PHARMACOLOGY AND EXPERIMENTAL THERAPEUTICS
Journal of the American Medical Association · 1909 · 7.9K citations · Full text
Keith T. Muir, Karin A. Kook, Craig Stern et al. · Journal of Chromatography B Biomedical Sciences and Applications · 1986 · 18 citations