Science and Technology Facilities Council · 2002 · 49 citations · 20 references
EngineeringNuclear EngineeringPhysicsInstrumentationNuclear PhysicsSpectroscopyCalibrationDetector SystemsNeutron DetectionRadiation InstrumentationRadiation SpectroscopyDetector Efficiency CalibrationInelastic Neutron ScatteringNeutron Detector CalibrationNeutron SpectroscopyNuclear SpallationSpectrometer CalibrationSpectrometer Design
The electron Volt Spectrometer (eVS) is an inverse geometry filter difference spectrometer that has been optimised to measure the single atom properties of condensed matter systems using a technique known as Neutron Compton Scattering (NCS) or Deep Inelastic Neutron Scattering (DINS). The spectrometer utilises the high flux of epithermal neutrons that are produced by the ISIS neutron spallation source enabling the direct measurement of atomic momentum distributions and ground state kinetic energies. In this paper the procedure that is used to calibrate the spectrometer is described. This includes details of the method used to determine detector positions and neutron flight path lengths as well as the determination of the instrument resolution. Examples of measurements on 3 different samples are shown, ZrH2, 4He and Sn which show the self-consistency of the calibration procedure.
20
The λ-Phenomenon of Liquid Helium and the Bose-Einstein Degeneracy
Nature · 1938 · 712 citations
On the Bose-Einstein Condensation
Physical Review · 1938 · 576 citations
Electron scattering from atoms, nuclei and nucleons
Geoffrey B. West · Physics Reports · 1975 · 303 citations
N.G. Sjöstrand · Nuclear Instruments and Methods in Physics Research · 1982 · 140 citations