IEEE Transactions on Biomedical Engineering · 1991 · 95 citations · 18 references
EngineeringBioengineeringBiomedical EngineeringBiomedical ImagingElectrophysiologyMedical PhysicsCardiac ImagingCardiovascular ImagingIntracardiac ElectrophysiologyCardiac Source LocalizationMagnetic Field-based LocalizationMagnetic Source LocalizationModel-based LocalizationVentricular Mapping
We describe a fast and numerically effective biomagnetic inverse solution using a moving dipole in a realistic homogeneous torso. We applied the localization model and high-resolution magnetocardiographic mapping to localize noninvasively the ventricular preexcitation site in ten patients suffering from Wolff-Parkinson-White syndrome. In all cases, the computed localization results were compared to the results obtained by invasive catheter technique. Using a standard-size torso model in all cases, the average 3-D distance between the computed noninvasive locations and the invasively obtained results was 2.8 +/- 1.4 cm. When the torso was rescaled to better match the true shape of the subject in five cases, the 3-D average was improved to 2.2 +/- 1.0 cm. This accuracy is very satisfactory, suggesting that the method would be clinically useful.
18
Numerical recipes: the art of scientific computing
W. H. Press, Brian P. Flannery, Saul A. Teukolsky et al. · Analytica Chimica Acta · 1987 · 3.3K citations
The Application of Electromagnetic Theory to Electrocardiology
A. Barnard, Ian Duck, M. Stuart Lynn et al. · Biophysical Journal · 1967 · 258 citations · Full text
Medicine, Cardiology, Physics +12