Journal of Biomedical Optics · 2010 · 59 citations · 10 references
EngineeringBioengineeringBiomedical EngineeringBiomedical ImagingBiomedical OpticsOpticsOptical CoherenceOptical Coherence TomographyOptical TomographyRefractive IndexTissue OpticsFocal LengthGradient-index LensesGradient-index OpticsOptical Coherence Tomography InstrumentationOptical Coherence Tomography ProbesOptimal DesignProbe Optimization
We report the numerical analysis of gradient index (GRIN) lens-based optical coherence tomography imaging probes to derive optimal design parameters. Long and short working distance probes with a small focal spot are considered. In each model, the working distance and beam waist are characterized and compared for different values of length and refractive index of the probe components. We also explore the influence of the outer tubing and refractive index of the sample media. Numerical results show that the adjustment of the maximum beam diameter and focusing angle at the end of the GRIN lens surface is very important for determining the optical performance parameters of the probe.
10
David Huang, Eric A. Swanson, Charles P. Lin et al. · Science · 1991 · 13.7K citations
In Vivo Endoscopic Optical Biopsy with Optical Coherence Tomography
Guillermo J. Tearney, Mark E. Brezinski, Brett E. Bouma et al. · Science · 1997 · 1.4K citations
Three-dimensional endomicroscopy using optical coherence tomography
Desmond C. Adler, Yu Chen, Robert Huber et al. · Nature Photonics · 2007 · 299 citations
Graded-index fiber lens proposed for ultrasmall probes used in biomedical imaging