AIChE Journal · 1958 · 35 citations · 7 references
EngineeringChemical EngineeringRheologyThermal AnalysisTransport PhenomenaCapillary RheologyFluid PropertiesHeat Transfer In FluidsSolution ChemistryViscous FlowBoilingBoiling PointCapillary ViscometryMedia ViscosityViscosity MeasurementViscosity-temperature Behavior
Abstract The viscosity of acetone‐water liquid solutions has been measured over the entire concentration range at temperatures from 20°C. to generally within 1° to 10°C. of the normal boiling point. A capillary suspended‐level viscometer was used, and the authors estimate that the results are accurate to ±0.2%. Both kinematic and absolute viscosities are given, and the results have been extrapolated to the boiling point of the solutions. The calibration of the Cannon‐Ubbelohde viscometer is discussed in detail.
7