Precise measurements with Bingham viscometers and Cannon master viscometers

J. F. Swindells, Robert C. Hardy, R. L. Cottington

Journal of research of the National Bureau of Standards · 1954 · 34 citations · 0 references

DOIFull text

Open access

Concepts

Abstract

A cri t ical study has been made of t he techniques used at the National Bureau of Standards wi th Bingham viscon~eters and Cannon kinemat ic vi~cometers. All corrections applicable to measurements wIth t hese lllstruments were cntlCally examined . Instruments of each type were calibrated using the viscosity of water at 20 C as the primary viscosity standard. The visco meters were used to determine the viscosit ies of four hydrocarbon liquids in t h e range 0.4 to 40 centipo ises. With each liquid the values obtained in the two types of viscometers were in agreement by 0.05 percent ~r better, indicating that no gross error was involved in the use of either instrument. It is considered however that t he inherent relat ive simplicity of operation of the kinematic viscometer n'1akes it a' preferable instrument for t his t.vpe of measurement.