IEEE Transactions on Nuclear Science · 1994 · 15 citations · 6 references
MedicineNuclear MedicineRadiologyDiagnostic ImagingDiagnostic Nuclear MedicineDiagnostic RadiologyMedical ImagingNuclear Medicine ImagingPositron Emission TomographyRadiation Imaging SystemsAcquisitionsMotion Correction In PetPatient Motion CorrectionPet Image ReconstructionPet QuantificationRadioactivity MeasurementTransmission Coefficient Measurement
PET transmission scanning in the presence of emission activity has been investigated on two scanners with orbiting pin transmission sources. One phantom consisting of varying sized spherical inserts in a uniform background and another containing simulated lungs and spine, a myocardium insert, and uniform background were imaged at various emission activity levels. Insert radioactivity concentration levels varied up to 3.5 /spl mu/Ci/mL. Transmission acquisitions were performed with the phantoms aligned with the emission scan position, and with the phantoms shifted to simulate patient motion between scans. Images were reconstructed with and without compensation for the emission contamination. Qualitative and quantitative evaluation of the resulting image sets was performed. The emission correction resulted in no visible artifacts and /spl sim/3% quantitative accuracy.< <ETX xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">></ETX>