Journal of Analytical Atomic Spectrometry · 1988 · 30 citations · 0 references
Natural SciencesChemical EngineeringChemistryAnalytical ChemistryElemental AnalysisSpectroscopyAnalytical MeasurementsAtomic Emission SpectroscopyElemental Composition AnalysisInstrumental AnalysisSample PreparationTrace Element AnalysisAtomic Emission SpectrometryAtomization ProcessesInductively Coupled Plasma Atomic Emission SpectroscopyInductively Coupled Plasma SpectrometrySample AtomizationSiliconSlurry PreparationViscosity
A simple and rapid method for the analysis of whole kaolin samples based on the nebulisation of slurries into an inductively coupled plasma for atomic emission spectrometry is described. The importance of nebuliser design and particle size on transport efficiency is discussed. In order to obtain acceptable results for slurry analysis it was necessary to use the silicon content of the kaolin as an internal standard element. This effectively corrected for variable atomisation efficiencies associated with different particle sizes. The value of dispersants, in this instance aqueous ammonia solution, in improving atomisation efficiency is demonstrated. The effect of viscosity variations arising from differing slurry concentrations (above 10%m/V) was corrected by the use of lanthanum as an internal standard.