VOLTAMMETRIC DETERMINATION OF 2-AMINOFLUORENE AND 2,7-DIAMINOFLUORENE USING CARBON PASTE ELECTRODE

Saulius Armalis, Natalija Novikova, Erika Kubilienė

Analytical Letters · 2002 · 12 citations · 7 references

Concepts

Abstract

ABSTRACT The voltammetric determination of 2-aminofluorene and 2,7-diaminofluorene based on their oxidation on carbon paste electrode is described. Britton-Robinson buffer and pH ranges 1–3 and 10–12 are recommended for the simultaneous determination of 2-aminofluorene and 2,7-diaminofluorene. The sensitivity of the determination can be enhanced substantially by adsorptive accumulation of analytes on carbon paste electrode in the pH range 10–12. The detection limits for the accumulation time 200 s were evaluated as 2.10−7 and 1.10−7 mol L−1 for 2-aminofluorene and 2,7-diaminofluorene, respectively. Relative standard deviations for the concentration range 1−10 × 10−7 mol L−1 do not exceed 10%. Keywords: 2-aminofluorene2,7-diaminofluoreneCarbon paste electrodeDifferential pulse voltammetryAdsorptive stripping voltammetry ACKNOWLEDGMENT JB thanks the Grant Agency of the Charles University (grant no. 253/2001/B-Ch/PrF). JZ thanks the Grant Agency of the Czech Republic (grant no. 203/01/1609). SA, NN and EK thank to the UNESCO participation project 00CZE408.