Magnetocardiography with a modular spin-exchange relaxation-free atomic magnetometer array

Robert Wyllie, M. Kauer, Gregory Smetana, Ronald T. Wakai, Thad Walker

Physics in Medicine and Biology · 2012 · 122 citations · 29 references

DOIFull text

Open access

Concepts

Abstract

We present a portable four-channel atomic magnetometer array operating in the spin-exchange relaxation-free regime. The magnetometer array has several design features intended to maximize its suitability for biomagnetic measurement, specifically foetal magnetocardiography, such as a compact modular design and fibre-coupled lasers. The modular design allows the independent positioning and orientation of each magnetometer. Using this array in a magnetically shielded room, we acquire adult magnetocadiograms. These measurements were taken with a 6-11 fT Hz(-1/2) single-channel baseline sensitivity that is consistent with the independently measured noise level of the magnetically shielded room.

References