Concepedia

Publication | Closed Access

Comparison of Retrospectively Self-Gated and Prospectively Triggered FLASH Sequences for Cine Imaging of the Aorta in Mice at 9.4 Tesla

21

Citations

19

References

2012

Year

Abstract

RSG-FLASH is preferable for cine imaging of the aorta. It provides the same quantitative data as PT-FLASH cine imaging but is less prone to flow and trigger artifacts. RSG-FLASH permits more homogeneous depiction of the cardiac cycle and is faster than the PT-FLASH sequence. PT-FLASH is more prone to misregistration of the respiratory cycle or the ECG by the external monitoring device used for acquisition. This effect may be even more pronounced in animals with disease models that are less stable in terms of heart and respiration rate during anesthesia.

References

YearCitations

Page 1