The liquid xenon detector for PET: recent results

V. Chepel, V. N. Solovov, J. van der Marel, I. Lopes, Paulo Crespo, L. Janeiro, D. Santos, R. Ferreira‐Marques, A. Policarpo

IEEE Transactions on Nuclear Science · 1999 · 30 citations · 10 references

Concepts

Abstract

A multiwire ionisation chamber filled with liquid xenon as a position sensitive detector for Positron Emission Tomography is studied. Both xenon scintillation, which provides a fast trigger, and ionisation charge for position and energy measurements are read. The detection efficiency of the detector is assessed in separate measurements of the probability of detection of the scintillation light and the charge signals due to 511 keV gamma-rays. These were measured to be about 1 and 0.82, respectively.