IEEE Transactions on Nuclear Science · 1999 · 30 citations · 10 references
EngineeringNuclear MedicineRadiologyNuclear InstrumentationRadiation DetectionRadiation MeasurementCharged Particle DetectionDetector TechnologyPositron Emission TomographyRadiation DetectorsRadiation Imaging DetectorsScintillation DetectorsPet Detector TechnologyPosition-sensitive DetectionPosition-sensitive DetectorsPositron Emission Tomography DetectionPositron Emission Tomography DetectorsRadiation Detection EfficiencyScintillationScintillation Detection
A multiwire ionisation chamber filled with liquid xenon as a position sensitive detector for Positron Emission Tomography is studied. Both xenon scintillation, which provides a fast trigger, and ionisation charge for position and energy measurements are read. The detection efficiency of the detector is assessed in separate measurements of the probability of detection of the scintillation light and the charge signals due to 511 keV gamma-rays. These were measured to be about 1 and 0.82, respectively.
10
Average energy expended per ion pair in liquid xenon
T. Takahashi, S. Konno, T. Hamada et al. · Physical review. A, General physics · 1975 · 138 citations
Liquid xenon scintillators for imaging of positron emitters
Louis Lavoie · Medical Physics · 1976 · 36 citations