Publication | Closed Access
400-Hz mechanical scanning optical delay line
126
Citations
7
References
1993
Year
Transient GratingPhotonicsEngineeringOptical PropertiesOptical TestingMechanical EngineeringPlanar MirrorTime MetrologyOptical Delay LineUltrafast OpticsOptical SystemsInstrumentationHigh SpeedTime-of-flight ImagingElectro-optics Device
We have developed a simple, high-speed, nearly vibration-free, mechanically scanned, optical delay line suitable for femtosecond time-resolved signal-averaging measurements. We demonstrate a 2-ps time window autocorrelator with a display updated at 400 Hz. The delay line uses a dithering planar mirror as a time-varying linear phase ramp in the spectral plane of a modified grating-lens femtosecond pulse shaper. The time delay is linearly proportional to the angular deviation of the mirror. The high speed and low vibration are a result of the extremely small angular changes required to generate a large time delay.
| Year | Citations | |
|---|---|---|
Page 1
Page 1