Computer‐controlled flow simulator for MR flow studies

Richard Frayne, David W. Holdsworth, Linda M. Gowman, Daniel W. Rickey, Maria Drangova, Aaron Fenster, Brian K. Rutt

Journal of Magnetic Resonance Imaging · 1992 · 77 citations · 21 references

Abstract

A novel computer-controlled flow simulator for use in magnetic resonance (MR) flow experiments was evaluated. The accuracy in constant-flow mode was better than 1%. The accuracy in pulsatile-flow mode was found to be dependent on the interconnecting tubing. The short-term and long-term reproducibilities of pulsatile waveforms were less than or equal to 0.4 mL/sec (1 standard deviation). Increased response times due to the lengths of tubing required in MR flow experiments were surmounted by using a modified tubing configuration and precompensated waveforms. Piston reversal was found not to cause major difficulties in MR flow experiments.

References

21