Natural SciencesBiomedical EngineeringRadiologyBiomedical ImagingDiagnostic ImagingDiagnostic RadiologyImage AcquisitionImagingMagnetic Resonance ImagingMagnetic Resonance Imaging PhysicsMedical ImagingMedical Imaging PhysicsArtifacts In Magnetic Resonance ImagingFourier TransformFourier Transform ImagingImage ArtifactsMagnetic ResonanceMagnetic Resonance PhysicsMotion ArtifactsPeriodic Motion
Artifacts from periodic motion during MR image acquisition have been studied. A mechanical device was built to oscillate a sample in a 0.15‑T imager, and 2‑D and 3‑D images were acquired at various frequencies and amplitudes, with simulations matching the experimental results. Movement locally blurs MR images and produces ghost artifacts along phase‑encoding directions.
Artifacts due to periodic motion during the acquisition of magnetic resonance (MR) images have been studied. A mechanical device was constructed to oscillate a small sample along any line within a 0.15-T Technicare imager. Two- and three-dimensional images were obtained using various frequencies and amplitudes of oscillation. Computer simulations of these experiments yielded images which agreed with the experiments. We demonstrated that movement influences MR images locally through blurring, and also generates ghost artifacts along the phase-encoding directions of the Fourier transform imaging technique.