Concepedia

Publication | Open Access

Confidence limits of arteriovenous fistula flow rate measured by the 'on-line' thermodilution technique

17

Citations

10

References

2003

Year

Abstract

A method to estimate CLs of flow rate measured by the thermodilution technique is presented and validated. Application demonstrates an accurate measurement of low flow, but limitations at higher flow and in detecting falls in flow. Appreciating the magnitude of such is critical to informed clinical decision making when using flow rate in an access surveillance programme.