IEEE Transactions on Medical Imaging · 2000 · 93 citations · 28 references
MedicineBiomedical EngineeringRadiologyBiomedical ImagingDiagnostic ImagingDiagnostic RadiologyImage ReconstructionMagnetic Resonance Image ReconstructionMagnetic Resonance ImagingMagnetic Resonance Imaging PhysicsMedical Image AnalysisMedical ImagingContinuous To Discrete MappingMagnetic ResonanceNon-cartesian ReconstructionReconstruction AlgorithmsSignal-to-noise RatioSignal-to-noise Ratio Enhancement
A minimum-norm least-squares image-reconstruction method for the reconstruction of magnetic resonance images from non-Cartesian sampled data is proposed. The method is based on a general formalism for continuous-to-discrete mapping and pseudoinverse calculation. It does not involve any regridding or interpolation of the data and therefore the methodology differs fundamentally from existing regridding-based methods. Moreover, the method uses a continuous representation of objects in the image domain instead of a discretized representation. Simulations and experiments show the possibilities of the method in both radial and spiral imaging. Simulations revealed that minimum-norm least-squares image reconstruction can result in a drastic decrease of artifacts compared with regridding-based reconstruction. Besides, both in vivo and phantom experiments showed that minimum-norm least-squares image reconstruction leads to contrast improvement and increased signal-to-noise ratio compared with image reconstruction based on regridding. As an appendix, an analytical calculation of the raw data corresponding to the well-known Shepp and Logan software head phantom is presented.
28
Numerical recipes in C: the art of scientific computing
Choice Reviews Online · 1993 · 18.1K citations · Full text
The Mathematics of Computerized Tomography
Frank Natterer, Ge Wang · Medical Physics · 2002 · 2.6K citations · Full text
Multi-planar image formation using NMR spin echoes
P. Mansfield · Journal of Physics C Solid State Physics · 1977 · 1.8K citations