On-Chip Contactless Four-Electrode Conductivity Detection for Capillary Electrophoresis Devices

F. Laugere, Rosanne M. Guijt, J. Bastemeijer, G. van der Steen, A. Berthold, Erik Baltussen, Pascalina Sarro, Gijs W. K. van Dedem, Michael J. Vellekoop, A. Bossche

Analytical Chemistry · 2002 · 111 citations · 20 references

DOIFull text

Open access

Concepts

Abstract

In this contribution, a capillary electrophoresis microdevice with an integrated on-chip contactless four-electrode conductivity detector is presented. A 6-cm-long, 70-microm-wide, and 20-microm-deep channel was etched in a glass substrate that was bonded to a second glass substrate in order to form a sealed channel. Four contactless electrodes (metal electrodes covered by 30-nm silicon carbide) were deposited and patterned on the second glass substrate for on-chip conductivity detection. Contactless conductivity detection was performed in either a two- or a four-electrode configuration. Experimental results confirmed the improved characteristics of the four-electrode configuration over the classical two-electrode detection setup. The four-electrode configuration allows for sensitive detection for varying carrier-electrolyte background conductivity without the need for adjustment of the measurement frequency. Reproducible electrophoretic separations of three inorganic cations (K+, Na+, Li+) and six organic acids are presented. Detection as low as 5 microM for potassium was demonstrated.