Concepedia

Publication | Closed Access

Corneal Thickness Measurements With the Topcon SP-2000P Specular Microscope and an Ultrasound Pachymeter

134

Citations

8

References

1999

Year

Abstract

The new noncontact Topcon specular microscope provides measurements of corneal thickness that are somewhat less than those of ultrasound pachymetry, but that seem to be more consistent from one operator to another, possibly as a result of the elimination of observer bias induced by probe placement required by the ultrasound unit. This consistency may be important in the comparison of measurements by different operators over time in patients being followed up after refractive surgery or other therapeutic interventions.

References

YearCitations

Page 1