Corneal thickness measurements with the Orbscan Topography System and ultrasonic pachymetry
250
Citations
14
References
1997
Year
In both studies, the Orbscan system obtained statistically significantly different and higher values for corneal thickness. Regression analysis suggests that over the range of values in this study, the two devices differ by a constant amount (intercept and slope). The nonzero intercept of this regression shows that the values from the devices differ and cannot be directly substituted for each other. We therefore conclude that in this study, Orbscan system measurements of corneal thickness were 23 to 28 microm greater than ultrasonic pachymeter measurements. Linear regression equations may be developed for the results of measurements from the two devices and used as a precise transformation factor for the values obtained with the two devices.