Magnetic Resonance in Medicine · 1986 · 110 citations · 8 references
EngineeringBioengineeringBiomedical EngineeringBiomedical ImagingDiagnostic ImagingMagnetic Resonance BiophysicsMagnetic Resonance ImagingMagnetic Resonance Imaging PhysicsMagnetic Resonance Signal FormationMagnetic Resonance SpectroscopyMedical ImagingMedical Imaging InstrumentationField Strength DependenceMagnetic Field DependenceMagnetic Field GradientsMagnetic Field InhomogeneityMagnetic Resonance Imaging InstrumentationRadiofrequency CoilsSignal-to-noise RatioSignal-to-noise Ratio Analysis
A method is proposed for measuring on the bench the NMR signal-to-noise ratio of rf probes, (over the range 1-100 MHz) and also the power deposited in patients during the imaging experiment. The technique is based on the principle of reciprocity, in that a direct relationship exists between the magnetic field generated (upon transmission) by a matched probe coil and the signal-to-noise ratio delivered by the same coil when used as a receiver. The construction and use of a calibrated sense coil for measuring the field is described, and the precautions and theory necessary for accurate measurement and understanding are outlined. Finally, the method is verified by comparison with a direct spectral measure of sensitivity obtained from a small doped water sample placed in NMR imaging equipment.
8