Radiation Protection Dosimetry · 2010 · 21 citations · 12 references
MedicineRadiologyDiagnostic ImagingRadiation DosimetryRadiation ProtectionBreast ImagingMammography PhysicsRadiation Dose MeasurementAverage Glandular DoseBreast DosimetryExposure MeasurementFull-field Digital MammographyMammographyMammography ImagingMammography Quality AssuranceSkin Exposure Assessment
This study was aimed at evaluating the mean glandular dose (MGD) and affecting factors during mammography examinations by a full-field digital mammography unit. An extensive quality control program was performed to assure that the unit is properly working. Required information including compressed breast thickness (CBT), breast parenchymal pattern and technical factors used for imaging were recorded. An entrance skin exposure measurement was also performed using slabs of polymethylmethacrylate with 2-8 cm thickness. On the basis of recorded information and measured data, the MGD was estimated for 1145 mammography examinations obtained from 298 patients. Mean CBTs of 4.9 and 5.8 cm and MGDs of 2 and 2.4 mGy were observed for craniocaudal and mediolateral oblique views, respectively. Significant correlation was seen between MGD and CBT, breast parenchymal pattern and applied kVp and mAs.