Publication | Open Access
Non-contact respiratory rate measurement validation for hospitalized patients
129
Citations
4
References
2009
Year
Unknown Venue
Respiratory Rate AccuracyMeasurementMedicineVentilationPatient SafetyWearable TechnologyRespiratory RatePatient MonitoringThoracic UltrasoundHospitalized PatientsDoppler RadarCardiologyEmergency MedicineHospital Medicine
This paper presents the first clinical results for validating the accuracy of respiratory rate obtained for hospitalized patients using a non-contact, low power 2.4 GHz Doppler radar system. Twenty-four patients were measured in this study. The respiratory rate accuracy was benchmarked against the respiratory rate obtained using Welch Allyn Propaq Encore model 242, the Embla Embletta system with Universal XactTrace respiratory effort sensor and Somnologica for Embletta software, and by counting chest excursions. The 95% limits of agreement between the Doppler radar and reference measurements fall within +/-5 breaths per minute.
| Year | Citations | |
|---|---|---|
Page 1
Page 1