IEEE Transactions on Instrumentation and Measurement · 1995 · 14 citations · 6 references
EngineeringElectrical EngineeringMetrologyMicroelectronicsUncertainty QuantificationDevice CharacterizationMeasurement StandardsJosephson Voltage StandardsReference Standard ComparabilityReference StandardsTransfer StandardsVoltage References
A comparison of the 10 V Josephson array voltage standards of the BNM/LCIE and of the BIPM has been carried out. An indirect comparison using Zener-diode based voltage references as transfer standards shows an agreement of 5 parts in 10/sup 9/ with an uncertainty (1/spl sigma/) of the same value. On-site indirect and direct comparisons show no difference between the two standards within the total uncertainty of 1.2 parts in 10/sup 10/.< <ETX xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">></ETX>
6
A 10-V Josephson voltage standard
C.A. Hamilton, F. L. Lloyd, Kao Chieh et al. · IEEE Transactions on Instrumentation and Measurement · 1989 · 79 citations