Applied and Environmental Microbiology · 1996 · 90 citations · 9 references
Life SciencesMicrobiologyAquatic MicrobiologyDiagnostic MicrobiologyMarine EcologyAquatic BacteriologyBioanalytical Method ValidationWater QualityWater SanitationEnterococciEnterococci MonitoringFalse Negative MeasurementMarine Bacterial EnumerationPathogen Detection In WaterRecreational Water MicrobiologyWaterborne Pathogen Detection
Enterolert (IDEXX Laboratories Inc., Westbrook, Maine), a semiautomated, most probable number method for enumeration of enterococci, was compared with the standard membrane filter method by parallel testing of 138 marine and freshwater recreational bathing water samples. No statistically significant difference and a strong linear correlation were found between methods. Culturing of 501 Enterolert test wells resulted in false-positive and false-negative rates of 5.1 and 0.4%, respectively. Less time for setup, incubation (24 versus 48 h), and reading of Enterolert permits more efficient monitoring of recreational bathing areas.