Measurement of Time-Quadrature Components of Microwave Signals

J. Richmond

IEEE Transactions on Microwave Theory and Techniques · 1955 · 19 citations · 2 references

Concepts

Abstract

A phase-sensitive coherent detector used for microwave laboratory measurements is described. The receiver measures the real (|E| cos alpha) and imaginary (|E| sin alpha) components of a signal E with equipment which is less elaborate than that required for measuring the amplitude |E| and phase alpha. Furthermore, many calculations are more convenient if E is presented in rectangular rather than polar form. Measurements made with the receiver on known fields in waveguides are included to demonstrate its accuracy. The receiver has a sensitivity of -125 dbw at 9,375 mc.