Magnetic Resonance in Medicine · 2003 · 174 citations · 17 references
MedicineBiomedical EngineeringRadiologyDiagnostic ImagingDiagnostic RadiologyMedical PhysicsDiagnostic UltrasonographyHepatobiliary ImagingLiver ImagingMagnetic Resonance ImagingMedical Imaging InstrumentationLaser AblationMultiple Baseline DesignPorcine ModelsRespiratory Motion CompensationRespiratory Motion Correction
A technique is presented for the acquisition of temperature maps in the presence of variable respiratory motion using the proton resonance frequency (PRF) shift. The technique uses respiratory triggering, diaphragm position determination with a navigator echo, and the collection of multiple baseline images to generate temperature maps. Laser ablations were performed in an ex vivo liver phantom undergoing variable simulated respiratory motion and in vivo in four porcine livers, demonstrating a reduction of artifacts in the computed temperature maps compared with conventional single baseline techniques, both uncorrected and corrected for motion.
17
Ultrasound quantitation of respiratory organ motion in the upper abdomen
Stephen Davies, Andrea Hill, Robert Holmes et al. · British Journal of Radiology · 1994 · 329 citations
MR imaging of laser-tissue interactions.
Ferenc A. Jólesz, A. Bleier, Peter D. Jakab et al. · Radiology · 1988 · 328 citations