Applied Physics B · 2010 · 10 citations · 30 references
Natural SciencesChemistryAnalytical ChemistrySpectroscopic AnalysisSpectroscopyAbsorption SpectroscopyInfrared SpectroscopyLaser SpectroscopyOptical SpectroscopyTime-resolved SpectroscopyUltrafast OpticsAtomic Line ShapesBandwidth AnalysisChemical SensingInstrumental Line BroadeningMolecular Absorption FeaturesSpectral Line BroadeningSpectral Line Shape AnalysisSpectral Line Shapes
Quantum cascade lasers (QCLs) have attracted considerable interest as an alternative tuneable narrow bandwidth light source in the mid-infrared spectral range for chemical sensing. Pulsed QCL spectrometers are often used with short laser pulses and a bias current ramp similar to diode laser spectroscopy. Artefacts in the recorded spectra such as disturbed line shapes or underestimated absorption coefficients have been reported. A detailed time-resolved high-bandwidth analysis of individual pulses during a laser sweep has been performed. Quantitative results for CH 4 absorption features around 1347 cm −1 (7.42 μm) fell short of the expected values for reasonable operating conditions of the QCL. The origin of the artefacts using short pulses was identified to be partly of the same nature as in the case of long laser pulses. A complex combination with the tuning principle was found, leading to an apparently increased instrumental broadening (effective line width) and underestimated concentrations at low-pressure conditions.
30
Jérôme Faist, Federico Capasso, Deborah L. Sivco et al. · Science · 1994 · 4.5K citations
The HITRAN 2004 molecular spectroscopic database