Improvements in the atomic-fluorescence determination of mercury by the cold-vapour technique

K. Clive Thompson, R. G. Godden

The Analyst · 1975 · 47 citations · 0 references

Concepts

Abstract

Relatively simple modifications to a cold-vapour, mercury-fluorescence detector have resulted in a large increase in sensitivity and have decreased the time required for each measurement. A 2σ detection limit of 0·02 ng was achieved.